Adjustment of display value error of hydraulic universal material testing machine

Adjustment of Display Value Error of Hydraulic Universal Material Testing Machine Adjustment of Display Value Error of Hydraulic Universal Material Testing Machine

The hydraulic universal material testing machine is a commonly encountered instrument in the mechanical metrology verification. The JJG139-1999 verification procedure requires that the allowable error of the indication value is usually not more than ±1%, and the metrology department carries out the measurement and verification, and the indication error is mainly It is reflected in the following ways:
First, the positive deviation of the indication value is out of tolerance.
Second, the indication value of the negative deviation is out of tolerance.
3. The indication error shows “before positive and negative” or “pre-negative and positive” on the dial and the individual points are out of tolerance.
In view of the above situation, the author believes that the adjustment is as follows:
The first case: after confirming that the installation level of the instrument meets the requirements of the regulations, first check whether the friction of the working part is too large. If yes, adjust the gap of the guide wheel to eliminate the friction and clean the working cylinder if necessary, such as eliminating friction. After the force is still affected, the tightening screw of the pendulum rod and the push plate connecting bushing in the force measuring part (reading mechanism) should be loosened, and the push plate should be adjusted inward. After tightening the tightening screw, The small plate is gradually calibrated and repeated several times until the test is qualified. If the small plate is qualified and the large and medium plates are still out of tolerance, the weight of B 铊 and C 应 should be increased appropriately until the test is qualified.
The second case: first check whether the force measuring piston is installed correctly, and whether the friction force is too large. After eliminating this cause, the push plate is adjusted to the outside, and the small plate is adjusted. If the large and medium plates are still out of tolerance, The weight of Bé“Š and Cé“Š should be appropriately reduced until the test is qualified.
The third case: the angle of the contact surface between the push plate and the tooth head should be changed at this time. The contact surface is usually fixed on the surface of the push plate by a flat piece of steel with two upper and lower screws. The adjustment method is to loosen the two screws. Open, use copper or other thin metal sheet on the upper side or the lower side between the steel sheet and the push plate (only on one side), in order to change the angle, repeatedly check and adjust until qualified.